CORiTEC splint comfort

CORiTEC splint comfort – wearing comfort redefined.
  • The ultimate wearing comfort: becomes more flexible at mouth temperature and adapts to the individual tooth situation
  • Combines the impossible: flexibility and strength
  • Highly biocompatible

CORiTEC splint comfort – Innovation Made in Germany!

Artikelnr: N/A Kategorier: ,

Beskrivning

HIGHLIGHTS

  • Occlusal splints
  • Reflex splints
  • Repositioning splints
  • Stabilization splints
  • Therapeutic splints

CORiTEC splint comfort is different from conventional splint materials. Thanks to its patented technology, CORiTEC splint comfort adapts to the individual situation in the mouth in a reversible way – every day anew. The material becomes much more flexible at mouth temperature, so that a pressure-free and unique wearing comfort is achieved. You forget that you are wearing a splint! CORiTEC splint comfort also has cushioning properties, which makes it ideal for patients with bruxism.

CORiTEC splint comfort is highly biocompatible. The flexibiliser used is the only one of its kind approved for medical devices. In contrast to other materials of this class, CORiTEC splint comfort contains no di-benzoyl peroxide, which might cause skin irritation.

Thanks to its high resistance to yellowing and plaque, CORiTEC splint comfort remains extremely aesthetic even after long use.

APPLICATION

  • CORiTEC splint comfort blanks are used for the CAD/CAM-production of occlusal splints. Use suitable cross-cut carbide burs only.
  • Do not clamp the blank too tightly in the milling machine, 0.6 Nm is sufficient.
  • For removing the splint from the blank use suitable cross-cut carbide burs or cutting discs suitable for composites. Do not apply pressure.
  • Minimum wall thickness: 0.9 mm (for aesthetic reasons 0,8 mm is permitted in the labial region)
  • In order to avoid plaque accumulation, subsequent polishing of the splints is essential.  Please pre-polish with suitable silicone polishers and goat’s hair polishers. Avoid heat development at all times.
  • Upon completion, use an ultrasonic bath with water at a maximum temperature of 35 °C. Cleaning solutions are not recommended.
  • For repair and addition of material please use THERMEO® powder liquid system by pro3dure medical GmbH to preserve the advantageous wearing comfort of the splint, which is lost when not using this system. Please observe the instructions for use.

Roughing:

  • 2 mm ball radius cutter – single cutter
  • Speed: 22.000 rpm
  • Radial path distance: 0,6 mm
  • Z depth increase: 0,6 mm
  • Feed: 1.600 mm/min

Finishing:

  • 2 mm ball radius cutter – single cutter
  • Speed: 22.000 rpm
  • Radial path distance: 0,1 mm
  • Z depth increase: 0,1 mm
  • Feed: 1.600 mm/min

Rest-roughing:

  • 1 mm ball radius cutter – single cutter
  • Speed: 28.000 rpm
  • Radial path distance: 0,1 mm
  • Z depth increase: 0,1 mm
  • Feed: 1.200 mm/min

The cooling/irrigation nozzles located on the spindle should be aligned with the tool tip. In this case, a uniform programming length must be observed or a tool length of 2 mm should be assumed. Air is sufficient as a cooling medium.

Ytterligare information

Vikt0,4 kg
Dimensioner14 × 12 × 4 cm
Storlek

16 mm, 20 mm